Technology Specification
Ventilator Mode:IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH,IMV, VCV, PCV
Display:High-definition 10.4” TFT colored LCD
Tidal Volume:20-2000 ml
Rate:1~99bpm
SIMV rate:2~20bpm
I:E:4:1~1:8
Trigger sensitivity:-1.0~2.0kPa
PEEP:0~25cmH2O
Pressure control:3~60cmH2O
SIGH: 1.5 times the inspiratory time (60 ~ 120 times adjustable)
Inspiratory platform:0~50%
Oxygen concentration :21%~100%
Pressure limit:5~60cmH2O
Parameters for ventilation monitoring
tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E,peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance, flow -volume loop, pressure-volume loop)
Safety alarm system
Airway pressure alarm
upper limit setting range 0.3~6.0 kPa
lower limit setting range 0.2~5.0 kPa
Per-minute ventilation volume alarm
upper limit setting range 3.0~30L/min
lower limit setting range 1.0~10L/min
VT alarm
upper limit setting range 10~2000ml
lower limit setting range 0~1800 ml
Sustained high-pressure alarm
it will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm
it will give sound and light alarm if there is no tidal volume input for 15 seconds
Power
AC 220V±10% 50Hz±1Hz UPS, Storage battery
Optional Configuration
Trolley,Air compressor
|