Flowmeter O₂ N₂O AIR Four Tube Notes: ▪ Flowmeter O₂ N₂O AIR Four Tube ▪ Flow range: O₂:(0.1-10)L/min; N₂O:(0.1-10)L/min; AIR:(0.1-12)L/min ▪ Rotating float ▪ Accuracy: ±10%FSL(p=1013hPa,T=20±3°C)(Each flow tube shall be calibated to ±10% of the indicated value for flow rates between 10% and 100% of full scale) ▪ Alarm function ▪ Oxygen Concentr-ation:21%-29% ▪ Leakage test: ≤5ml/min@200kPa ▪ Oxygen filling fast ▪ N₂O AIR switching function ▪ When oxygen is below a certain value, there is a cut-off function
|